NEW! Dental City Development Website

HotSpot Infrared Thermometer

SKU
DC00019977
   
  • Description
  • Manufacturer Special
  • SDS Sheets
  • ICI’s HotSpot Non-contact Infrared Thermometer ensures accurate and stable body temperature readings. It’s effective for adults as well as newborn babies and young children with high mobility tendencies.

    Readings can be taken as close as 5 cm or as far as 15 cm from target zone. Being noncontact makes for a more sanitary environment and reduces the chance to spread infectious diseases.

    Temperature shows on the LCD screen. Users can set an alarm for readings considered abnormally high. Onboard data logger stores last 32 readings. The HotSpot IR is fits in lab coats and bags for portability.

    Features include:

    • Precise non-contact temperature measurements
    • Select between °C or °F readout
    • Select between Body and Surface temperature
    • Alarm sound for threshold
    • Memorization of the last 32 measurements
    • Automatic selection range
    • Display resolution 1/10th of a degree
    • Backlight LCD display
    • Auto power off

     Safety Conformance:

    • ASTM E1965-1998
    • EN 980: Graphical symbols for use in the labeling of medical devices
    • EN 1041:Information supplied by the manufacturer with medical devices
    • EN 60601-1: Medical electrical equipment Part 1: General requirements for safety (IEC: 60601-1:1998)
    • EN 60601-1-2: Medical electrical equipment Part 1-2: General requirements for safety Collateral standard

    Range:

    • Body Mode: 32.0 °C to 42.5 °C (89.6 °F to 108.5 °F )
    • Surface Mode: 0 °C to 60 °C (32 °F to 140 °F)
    • Resolution: ±1 °C
    • Basic Accuracy: ASTM E1965-1998 (2003)
    • 32 °C to 35.9 °C (93.2 °F to 96.6 °F) with (± 0.3 °C/ ± 0.5 °F)
    • 36 °C to 39 °C (96.8 °F to 96.6 °F to 102.2 °F) with (± 0.2 °C/± 0.4 °F)
    • 39 °C to 42.5 °C (102.2 °F to 108.5 °F) with (± 0.3 °C /± 0.5 °F)
    • Measuring Distance: 5 cm - 15 cm
    • Response Time: 0.5 Seconds